Determination of Tadalafil in pharmaceutical preparation of combination therapy with complex of vitamins and minerals by HPLC-UV detector

The method of determining Tadalafil in a mixture with a complex of vitamins and minerals by HPLC was verified to be linear, precise, selective and accurate with the use of WATERS Nova-Pack® C8 (3.9x150) mm; 4 ?m column. The mobile phase used was water:acetonitrile:trifluoroacetic acid (60:40:0.1) at...

Full description

Saved in:
Bibliographic Details
Main Authors: Muniz, Ana, Rosa, Paulo
Format: Online
Language:Portuguese
Published: Universidade Estadual de Campinas 2019
Subjects:
Online Access:https://econtents.sbu.unicamp.br/eventos/index.php/pibic/article/view/2048
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:The method of determining Tadalafil in a mixture with a complex of vitamins and minerals by HPLC was verified to be linear, precise, selective and accurate with the use of WATERS Nova-Pack® C8 (3.9x150) mm; 4 ?m column. The mobile phase used was water:acetonitrile:trifluoroacetic acid (60:40:0.1) at the flow rate of 1 mL/min with UV detection at 285 nm at ambient temperature. Linearity was determined between 0.08 and 0.11 mg/mL with a correlation factor of 0.998. Standard deviations of performed tests of precision and accuracy were insignificant and components of the matrix did not interfere in the retention time of Tadalafil, the method is selective.
ISSN:2596-1969